What Accuracy Percentage Plus Digits Means on a Multimeter

As an Amazon Associate we earn from qualifying purchases.

An accuracy spec written as a percentage plus digits, such as “±(0.5% + 10)”, gives the largest error a multimeter reading can have under stated conditions. The percentage part scales with the value you measure. The digits part adds a fixed number of last-place steps on the display for the range you selected. Add the two together and you get a band around the displayed value. The person typing “accuracy percentage plus digits multimeter” wants to know how far a displayed number can be from the true value, and how to work out that band for a specific reading.

Quick answer

  • The percentage is a share of the reading you are measuring, so a larger reading gets a larger allowance.
  • The digits (also called counts) are steps of the last displayed digit for the range in use. They depend on resolution, not on the reading.
  • To find the band, take the percentage of the reading, convert the digit count into the value of one last-place step, add the two, and subtract or add that total from the displayed value.
  • The spec sets a worst case. It does not promise that every reading lands at the middle of the band.

How we chose

We chose two explainers that describe this spec format directly. The first is a technical note from B&K Precision on reading accuracy specifications. The second is a Fluke learning article on accuracy and precision in digital multimeters. We read both pages in full and used only the definitions and worked examples they state. We did not use search snippets, and we did not take numbers from any product listing. The examples below are calculations based on the formula each source gives, and we show the arithmetic so you can check it.

What “percent of reading” means

The B&K Precision note describes accuracy as a value often written as percent of reading plus or minus counts or digits. The percentage is applied to the reading you are looking at, not to the full scale of the range. That is the key difference from a spec written as percent of range. On a meter that is accurate to a percentage of reading, a small reading has a small error allowance, and a large reading has a larger one. The Fluke article defines accuracy as the largest allowable error under specific operating conditions, expressed as a percentage, and it stresses that the conditions matter as much as the number.

Because the percentage moves with the reading, you cannot read one accuracy figure off the front of a meter and assume it applies everywhere. Work out the band for the value you actually measure.

What “digits” or “counts” means

The digits part of the spec is tied to resolution. Resolution is the smallest step a meter can show on a given range. The B&K Precision note says the counts are added to the end of the displayed number, based on the resolution of the reading. The Fluke article explains that counts describe how many steps a meter can display, and that counts are now the more common way to classify a meter. It gives a simple example: a 1999-count meter cannot show a tenth of a volt at 200 V or more, because its display runs out of steps before that point.

The Fluke article also says that a 3½-digit meter shows three full digits plus a half digit, while a 4½-digit meter has four full digits plus a half digit and so has finer resolution. Fluke’s own range and resolution table lists 3.000 V with 1 mV resolution, 30.00 V with 10 mV, and 300.0 V with 100 mV. A digits term in the spec therefore changes with the range you select, even when the percentage stays the same.

One point the sources make clearly: resolution is not the same as accuracy. A meter can display very fine steps and still have a wide error band, and a meter with coarse steps can be more accurate than its display suggests. Read both parts of the spec.

Worked example 1: a 100 V reading with ±2 percent

Take a reading of 100.0 V on a meter with a spec of ±2 percent of reading. Two percent of 100.0 V is 2.0 V. The band runs from 98.0 V to 102.0 V. The Fluke article gives the same result for this case.

Now add a digit term. The Fluke article gives a ±(2% + 2) example for the same 100.0 V reading, with a band from 97.8 V to 102.2 V. The 2 counts here are 0.2 V, because the display shows one decimal place at that range. Two percent is 2.0 V and two counts are 0.2 V, so the total allowance is 2.2 V.

Worked example 2: a 2.0000 V reading with ±(0.5% + 10)

The B&K Precision note uses a 5 V DC range on a 50,000-count meter with a spec of ±0.5% plus 10 counts. The measured reading is 2.0000 V. Half a percent of 2.0000 V is 0.0100 V. Ten counts on a 50,000-count display at this range is 0.0010 V. The total allowance is 0.0110 V, so the band runs from 1.9890 V to 2.0110 V. The note shows the same numbers.

The note also shows what happens at lower resolution. On a display with a 2.000 V resolution, the same spec gives a band from 2.000 V to 2.020 V for this reading. The percentage did not change, but the counts did, so the band changed too. This is why the range you pick affects the accuracy band you get.

A quick calculation table

The table below uses the formula from the B&K Precision example. Each row is a calculation, not a measured result.

Displayed reading Percent part (0.5%) Digit part (10 counts at 0.0001 V) Band
2.0000 V 0.0100 V 0.0010 V 1.9890 V to 2.0110 V
1.0000 V 0.0050 V 0.0010 V 0.9940 V to 1.0060 V

At a lower reading, the percentage allowance shrinks, but the digit part stays the same. For very small readings, the digit part can become the larger share of the error, which is a common reason low readings look less certain than large ones.

Reading vs. range

Some specs are written as a percentage of range instead of reading. If your meter’s spec says percent of range, the error allowance is the same across the range regardless of the value shown. The difference matters at small readings. Check the wording in the spec table before you calculate anything. Both B&K Precision and Fluke describe percent of reading as the common form for digital meters, but a specific meter’s manual decides which applies to it.

Conditions in the spec

The Fluke article notes that accuracy is stated for specific operating conditions. A spec may apply only within a temperature band, for a set time after calibration, or within a stated humidity range. Outside those conditions, the error can be larger. The sources we opened describe this principle but do not list typical temperature coefficients, so this article does not give one.

The Fluke article also notes that the required accuracy depends on the job. Its example is a receptacle check, where a meter at ±3 percent is described as adequate for asking whether a standard receptacle is energized, because normal receptacle voltage can vary by about 5 percent or more. That example is about the meter’s fit to a task. It is not a procedure, and this article does not describe how to test a receptacle.

What the spec does not tell you

The band tells you how far the reading can be from the true value under the stated conditions. It does not tell you that the meter is calibrated today, that the leads are in good condition, or that the circuit you are measuring is stable. A reading that drifts or jumps on a noisy circuit can fall outside the band even on a good meter, because the band describes the meter, not the signal.

Keep one practical rule in mind. If you need a result to a specific tolerance, compare the calculated band against that tolerance before you decide whether the meter is good enough. A meter whose band is wider than your tolerance cannot prove the result, even when the display looks precise.

For a related explainer on how current clamps sense current, see our article on how a current clamp works. For a clamp-based non-contact check and its limits, see our guide to Klein non-contact voltage detectors. Both articles use the same resolution and accuracy ideas in a different context.

What we could not verify

  • The two sources we opened do not give temperature coefficients or calibration intervals for any named meter, so this article does not state them.
  • We did not open a manufacturer datasheet for a specific multimeter model, so the bands above are worked examples from the formula in the sources, not the spec of any product.
  • The sources do not say how a particular meter rounds its last digit. Where rounding rules matter, check the manual.

FAQ

Is a higher digit count always more accurate?

No. A higher digit count gives finer resolution, which means smaller display steps. Accuracy is a separate number in the spec. A meter with more digits can still carry a wider accuracy band than a meter with fewer digits.

Does the percentage part or the digits part matter more?

It depends on the reading. At large readings the percentage usually sets most of the band. At small readings on a fine range, the digits part can become a larger share of the allowance. Calculate both parts for your actual reading.

Why does the same meter give a different band on different ranges?

The percentage stays the same, but the value of one last-place step changes with the range. Fluke’s range table shows different resolutions for 3 V, 30 V and 300 V ranges, so the digit term changes with the range you select.

Can I use the band to decide if a reading is wrong?

Only partly. The band tells you the allowed error for a correctly calibrated meter under the stated conditions. A reading outside the band suggests a problem with the meter, the conditions, or the signal, but the spec alone does not tell you which one.

Bottom line

Read the percentage as a share of the reading and the digits as steps of the last displayed digit on the range you chose. Add the two to get the band, and compare that band with the tolerance your job needs before you trust the number.

Sources

  • B&K Precision, How to Read Accuracy Specifications: https://www.bkprecision.com/knowledge/technical-notes/5/how-to-read-accuracy-specifications (opened 2026-10-30)
  • Fluke, Why Digital Multimeter Accuracy and Precision Matter: https://www.fluke.com/en-ca/learn/blog/digital-multimeters/accuracy-precision (opened 2026-10-30)