Hp 1740a Oscilloscope

Hp 1740a Oscilloscope

As an Amazon Associate we earn from qualifying purchases. This guide contains no paid product placements.

The HP 1740A oscilloscope is a 100 MHz, dual-channel analog scope with a delayed time base, made by Hewlett-Packard in the 1970s and later sold under the Agilent and Keysight names. It is a bench instrument for looking at waveforms, not a meter, and it is still found on used-equipment shelves. This guide explains the HP 1740A oscilloscope in plain terms: its bandwidth, rise time, deflection and sweep ranges, input limits and triggering, using the 1976 operating and service manual as the main reference.

Quick answer

The 1740A is a 100 MHz analog scope with a rise time of under 3.5 ns. Its vertical range runs from 5 mV/div to 20 V/div, and its main sweep runs from 50 ns/div to 2 s/div. Its maximum DC input is 250 V in 1 M ohm mode, so it is not a substitute for a high-voltage probe. Note that the manual gives its own sweep figures, and a retailer listing gives slightly different ones, so use the manual for detailed work.

How we chose these sources

Our main source is the HP 1740A Operating and Service Manual, part number 01740-90909, printed in August 1976, which we opened as a PDF from the Bitsavers archive. We read its specification table directly. For a second, independent check we opened a used-equipment retailer page, which lists some specifications. We did not open the DIYAudio forum thread or the ManualsLib copies, so those are not used here.

Bandwidth and rise time

The manual defines bandwidth as the 3 dB point below a 6-division reference signal. DC-coupled operation runs to 100 MHz in both the 50 ohm and 1 M ohm input modes. In AC coupling, the lower edge is about 10 Hz, and about 1 Hz when 10:1 probes are used. A bandwidth limit switch cuts the upper edge to about 20 MHz, which helps reduce noise on slow signals. The rise time is under 3.5 ns, measured from the 10 percent to 90 percent points of a 6-division step.

The retailer page we opened also lists 100 MHz bandwidth and gives the 5 mV/div sensitivity, plus 1 mV/div with the X5 vertical magnifier. It does not list rise time or input values.

Vertical deflection and input

  • Deflection factor: 12 calibrated steps from 5 mV/div to 20 V/div in a 1-2-5 sequence, accurate within 3 percent. The vernier extends the maximum to at least 50 V/div. An UNCAL light shows when the vernier is off its calibrated position.
  • Magnifier (X5): increases each deflection factor by five, with a maximum sensitivity of 1 mV/div on channels A and B.
  • Input impedance: selectable, either 1 M ohm plus or minus 2 percent in parallel with about 20 pF, or 50 ohm with a stated SWR of 1.4 or lower at 100 MHz.
  • Maximum input: 250 V (DC plus peak AC) or 500 V peak-to-peak AC at 1 kHz or less in 1 M ohm mode. In 50 ohm mode the maximum is 5 V rms, so do not apply a voltage source to the 50 ohm input that could exceed that.
  • Display modes: channel A, channel B, both channels alternate or chopped (about 250 kHz chop rate), algebraic A plus B, and a trigger-view mode that shows the trigger signal alongside the channels.
  • Differential mode: CMRR is at least 20 dB from DC to 20 MHz, measured with a common-mode signal of up to 8 divisions.

Horizontal sweep and accuracy

The main time base has 24 ranges from 50 ns/div to 2 s/div in a 1-2-5 sequence. The delayed time base has 18 ranges from 50 ns/div to 20 ms/div. A 10X magnifier expands every sweep by ten and extends the fastest sweep to 5 ns/div. Main sweep accuracy is plus or minus 3 percent at 15 to 35 degrees C, widening to plus or minus 4 percent at 0 to 15 degrees C and 35 to 55 degrees C. The manual adds 1 percent to the 50 ms/div to 2 s/div ranges.

The delay is calibrated from 0.5 to 10 times the main sweep setting, and a dual-time-base mixed mode lets the main sweep drive the first part of the trace while the faster delayed sweep completes it. The retailer page does not list sweep ranges, so the sweep figures here come from the manual alone.

Triggering

The trigger source can be channel A, channel B, composite of the displayed signal, or line frequency. Coupling options include AC, DC, low-frequency reject and high-frequency reject, with an adjustable trigger holdoff. External trigger inputs have about 1 M ohm and 20 pF input, and a maximum external input of 250 V DC plus peak AC or 500 V peak-to-peak at 1 kHz or less. A single-sweep mode arms the sweep for one pass and lights an indicator when it fires.

Accessories and what came in the box

The manual’s accessories list includes two 10:1 divider probes (HP model 10006D) and a storage pouch. Probes matter here because a 10:1 probe extends the usable input range and changes the low-frequency response, so check the probe’s rating before connecting it to a mains-related signal.

Safety notes

The 1740A is a line-powered instrument with internal high voltage. Do not open the case with the power cord connected, and use only a properly rated probe. The 250 V maximum input is a limit on the front end, not a safe working voltage for line-connected work. Keep the scope’s ground connection on the reference point of the circuit, because a floating or wrong ground can damage the instrument or the device under test.

Who the 1740A suits

The 1740A suits a hobbyist or repair technician who wants a classic analog scope for audio, power supply ripple and general signal checks up to 100 MHz. It has no digital storage or readout, so measurements come from reading the graticule. If you need a handheld meter for voltage, current and resistance checks rather than waveforms, see our guide to the best hand-held digital multimeters. For a simple voltage reading on a live circuit, our guide to the best digital volt meters covers that job.

What we could not verify

We could not verify the 1740A’s performance on a bench, because we had no unit to test. The service manual is a 1976 document, so its figures describe the instrument as designed then, and we could not confirm that a used unit still meets them. The retailer page we opened gives only the bandwidth and sensitivity, and it does not give rise time, input or sweep figures, so those rest on the manual alone. We did not confirm which serial-number revision of the manual applies to a given unit.

FAQ

What is the bandwidth of the HP 1740A?

The manual gives DC to 100 MHz in both 50 ohm and 1 M ohm modes, with a bandwidth limit that cuts the upper edge to about 20 MHz.

What is the rise time?

Under 3.5 ns, measured from 10 percent to 90 percent of a 6-division step.

What is the maximum input voltage?

250 V DC plus peak AC, or 500 V peak-to-peak AC at 1 kHz or less, in 1 M ohm mode. In 50 ohm mode the limit is 5 V rms.

Does the 1740A store waveforms?

No. It is an analog instrument with a CRT display and no digital storage.

Where do the sweep figures come from?

Only from the 1976 manual. The retailer page we opened does not list sweep ranges, so check a unit’s front panel before relying on them.

Sources

  • Hewlett-Packard, 1740A Oscilloscope Operating and Service Manual, part number 01740-90909, August 1976, via Bitsavers: https://ftp.mirrorservice.org/sites/www.bitsavers.org/test_equipment/hp/oscilloscope/01740-90909_1740A_Oscilloscope_Operating_and_Service_Manual_197608.pdf (opened 2026-10-29)
  • TestWorld, Keysight 1740A 100 MHz 2-channel oscilloscope product page: https://testworld.com/product/keysight-1740a-100mhz-2ch-oscilloscope/ (opened 2026-10-29)