How to Zero Out Lead Resistance on a Multimeter
As an Amazon Associate we earn from qualifying purchases.
To zero out lead resistance on a multimeter, set the meter to the ohms range, touch the two test-lead tips together, and press the REL (relative) key so the meter stores the leads’ own resistance as the reference. From then on, the display shows the resistance of the component above or below that stored value. Fluke’s resistance guide and its support article both describe the leads as adding a small resistance, typically 0.2 to 0.5 ohms, which matters when you measure very low resistances such as wire runs, fuses or switch contacts.
Quick answer
- Set the meter to ohms, insert the leads in the COM and VΩ jacks, and touch the tips together.
- Read the small value the leads show. Press REL so the display reads zero.
- Measure the component. The meter subtracts the stored lead value from the reading.
- If the leads show more than 0.5 ohms when shorted, the Fluke support article says to replace them.
The person typing this keyword wants
The person typing “how to zero out lead resistance on a multimeter” wants a repeatable way to remove the resistance of the test leads from a low-resistance reading. They are usually checking a short run of wire, a jumper, a fuse, a switch contact or a connection, where a few tenths of an ohm changes the answer. This guide gives the steps in order, the numbers the sources give, and the points where the sources stop.
How we chose
We limited this guide to two sources, both from Fluke, that we opened on 2026-10-30. The first is Fluke’s learning article “How to Measure Resistance with a Digital Multimeter”. The second is Fluke support article KA-02625, which covers the lead-resistance limit and the REL function. We did not use search snippets, forum posts or product-listing text for any claim. Where the two sources describe the same point, we state it once. We did not pick numbers from one source over another, because the sources agreed on the values that matter here.
The choice of a single manufacturer’s sources has a cost. The sources describe Fluke meters, and they do not compare REL behavior across other brands. We say so in the sections below rather than filling that gap from memory.
Why test leads add resistance
Every test lead is a conductor, and it has a small resistance of its own. The Fluke resistance guide states that the leads add a small resistance to the reading, and gives a typical range of 0.2 to 0.5 ohms. When the component being measured is itself only a few tenths of an ohm, the lead value is the same order of size as the reading, which is why the Fluke guide ties the relative mode to very low resistances. A reading of a few tenths of an ohm on a shorted pair of leads is not, by itself, a fault.
The support article KA-02625 says the same thing in practical terms. A reading above 0.0 ohms when the leads are shorted together is normal, because the leads have an inherent resistance. The article does not treat that small value as a meter problem.
What you need before you start
- A digital multimeter with an ohms function and a REL or relative mode. The support article names the Fluke 87V as a meter with REL. The sources we opened do not list every model that has the function.
- Test leads in good condition, with intact insulation and clean tips.
- The component out of the circuit, or at least isolated, for the best reading.
Step 1: de-energize and discharge first
Fluke’s resistance guide says to turn off power and discharge any capacitors before you measure resistance. If the work touches mains wiring, de-energize the circuit at its breaker, lock it out and tag it, and confirm that it is dead with a suitable tester before you connect the leads. This guide does not include steps for probing a live outlet or an energized panel.
Step 2: set the meter and the leads
Set the rotary switch to the ohms position. The Fluke guide describes inserting the leads in the COM and VΩ jacks. It also describes fixed range selection as an option. For a very low resistance, choose a range where the meter can resolve a small value, and use the relative mode to remove the lead value from the reading.
Step 3: short the leads and read the lead value
Touch the two probe tips firmly together and let the display settle. The Fluke guide says that touching the leads together should show 0 ohms in an ideal case, and support article KA-02625 says that a small value above 0.0 ohms is normal. Write down the number. It is the lead resistance you will remove in the next step. Keep the tips clean and dry while you do this. The Fluke guide lists dirt, solder flux, oil and body contact with the metal tips as causes of unstable readings.
Step 4: press REL to store the reference
With the leads still shorted, press REL. The Fluke guide describes the meter storing the present value as the reference, and the display then reads zero. From this point the meter shows the difference between each new reading and the stored value. The guide says that relative mode automatically subtracts the test lead resistance. Support article KA-02625 notes that meters such as the Fluke 87V use the REL function to store the lead resistance so that the meter can compensate for it.
Step 5: measure the component
Place the probe tips across the component. Read the display. The value shown is the component resistance after the stored lead value has been removed. Remove the leads from the component before you change any settings or swap leads, so that the stored reference is not mixed with a new circuit.
Reading the numbers: a worked table
The table below sets out the values the two sources give, and what each one means for your zeroing step.
| Shorted-lead reading | What the sources say | What to do |
|---|---|---|
| 0.0 ohms | The Fluke guide describes this as the ideal display with the leads shorted. | Press REL if you want the display to read zero during later measurements. |
| 0.2 to 0.5 ohms | The Fluke guide gives this as the typical lead resistance. | Zero the leads with REL before a low-resistance reading. |
| Above 0.5 ohms | KA-02625 says the leads should measure 0.5 ohms or less, and that higher values call for replacement. | Replace the leads before you rely on low-resistance readings. |
The arithmetic is simple. If the leads show 0.3 ohms when shorted, you press REL, and the meter stores 0.3 ohms as the reference. If the component then reads 0.8 ohms on the display, the value shown after the reference is applied is 0.5 ohms. That is a worked example from the numbers above, not a measurement of any particular product.
Where zeroing does not fix the reading
Zeroing removes the lead value, but it does not correct every error. The Fluke guide notes that parallel paths can lower a resistance reading, and it recommends removing the component from the circuit for the best result. It also advises checking the schematic for parallel paths. A reference stored with the leads shorted cannot tell the meter about a second conductor in the circuit. Dirty contacts and poor connections also raise the reading, and the zero does not remove them.
A second point: REL stores a value for one pair of leads and one measurement setup. The support article does not describe how long the stored reference stays valid, or what happens if the range changes. Check your meter’s manual for those details, and zero again whenever you swap leads.
Which meters have REL
The sources we opened name REL on the Fluke 87V in the support article, and they describe the relative mode in the Fluke resistance guide. They do not say which other Fluke models have the function, and they do not cover other brands. If your meter has no REL key, the same idea is to read the lead value by hand while the leads are shorted, and subtract that value from each later reading. The support article does not describe that method, so we have not presented it as a Fluke procedure. Treat it as your own arithmetic and check it against the manual.
Common mistakes
- Zeroing with the leads in a different jack or range than the one you will use for the measurement.
- Holding the metal tips while reading. The Fluke guide lists body contact among the causes of unstable readings.
- Measuring a component still connected to a circuit, which can add parallel paths.
- Assuming a small lead value is a fault. The support article says it is normal up to the limit.
- Using worn leads. The support article says leads above 0.5 ohms should be replaced.
How the leads fit into a continuity check
For a broader look at voltage and resistance basics on the same meter, see our guide to using a voltmeter. A continuity check uses the same leads and the same meter, but it reads a tone rather than a resistance value. If you are only checking whether a path is closed, use the continuity function described in our guide to how to use a continuity tester. Zeroing matters when you need the actual resistance of a low-value path, not just a yes or no answer.
What we could not verify
- Whether two-wire and four-wire measurement change the lead correction. The Fluke guide we opened does not discuss that comparison, so this guide does not make a claim about it.
- The lead resistance of any particular lead set. The 0.2 to 0.5 ohm figure is a typical range from the Fluke guide, not a measurement of your leads.
- Which non-Fluke meters have a relative mode, and how each one stores the reference.
- How long a stored REL value holds after the meter is switched off, or after the range changes.
- Any specific accuracy figure for low-resistance readings on any meter.
FAQ
Do I need to zero the leads before every measurement?
The sources we opened do not give a fixed interval. Zero the leads whenever you swap them, and whenever a low-resistance reading matters. Check your meter’s manual for any power-off behavior.
Is a 0.3 ohm reading on shorted leads a sign the meter is broken?
No, according to support article KA-02625, a reading above 0.0 ohms when the leads are shorted is normal. The article says to replace the leads only when they measure above 0.5 ohms.
Does zeroing fix a damaged lead?
No. Zeroing removes the stored lead value, but a lead that measures above the 0.5 ohm limit should still be replaced, per the support article.
Can I zero the leads for voltage readings?
The sources we opened describe REL for resistance measurements only. They do not describe a zero function for voltage readings.
Why does the reading drop when I measure a wire still in a panel?
The Fluke guide notes that parallel paths can lower a resistance reading. Remove the component from the circuit when you can, and check the schematic for parallel routes.
Bottom line
Short the leads, read the small value, press REL, and then measure. Replace any lead that shows more than 0.5 ohms when shorted, and take every low-resistance reading with the component out of the circuit.